Bernoulli's Theorem Demonstration module sits on a hydraulics bench and tests Bernoulli's equation along a convergent-divergent duct. An 8-tube manometer reading 0 to 300 mm of water shows the static head at each tapping, and a Pitot tube moved along the conduit gives the head in each section. Removable conduit ends let the duct face the flow either way.
| Parameter | Detail |
|---|---|
| Narrowing angles | 21° downstream, 10° upstream |
| Manometer | 8 tubes, 0 to 300 mm of water, all pressure taps connected, with water collector |
| Throat diameter | 25 mm |
| Conduit | Removable ends; fits in convergent or divergent form relative to the flow |
| Probe | Pitot tube moving along the conduit, measuring the head in every section |
| Flow control and connections | Control valve at the module end; inlet female coupling to the bench supply; outlet hose to the volumetric measuring tank |
| Mounting | Hydraulics bench top, adjustable legs for levelling |
| Construction | Anodized aluminium / MS powder-coated structure; painted steel panel |
| Size and weight (approx.) | 800 x 450 x 700 mm; 15 kg (33 lb) |
| Product code | SLE/MCE/33 |
| Minimum order | 10 units |
| Dispatch | Usually within 7 to 10 working days |
| Manufacturer | ScienceLab (Jain Scientific Equipments Pvt Ltd), Ambala, India |
Setting up and purging. Level the module on the bench top with its adjustable legs and couple the inlet to the bench supply. Run water with the control valve partly open until every manometer tube has filled and no bubbles remain in the tapping lines. Then set the valve so all levels sit on the scale.
Convergent duct readings. With the duct convergent to the flow, set a steady flow and read all 8 tubes once the levels stop moving. Time the filling of the bench measuring tank with a digital stopwatch to get the flow rate, then divide it by the section area at each tapping to find the velocity.
Pitot traverse. Move the Pitot tube along the conduit and note the head it shows beside each tapping. It reads the total head at its tip, so subtracting the static head from the matching manometer tube leaves the velocity head. Plot static, velocity and total head against distance along the duct.
Divergent duct and losses. With the ends fitted as labelled, the water narrows through the 10° upstream taper and widens through the steeper 21° downstream taper. Reverse the conduit ends and repeat at the same flow rate, so the gentler 10° taper becomes the widening one. Compare how much of the pressure drop is recovered after the throat in each arrangement; the shortfall shows energy lost to friction and turbulence. The friction loss in pipe apparatus measures such losses directly, and the orifice and Venturi meter demonstrator turns the same pressure drop into a flow measurement.
A bubble in a tapping line or tube adds a false column and makes that level wrong or unsteady. Run water through the duct until all tubes fill and the lines are clear, and tap the tubes lightly to free bubbles that cling. Readings taken with air present will not form a smooth curve.
At the throat, where the cross-section is smallest. The same flow must pass every section, so the water speeds up as the area falls and the static head drops to its minimum there. It rises again as the duct widens and the water slows down.
The equation assumes steady, frictionless flow of a fluid that cannot be compressed. Real water loses some energy to wall friction and turbulence, so the total head falls slightly from inlet to outlet. The drop is usually larger after the throat, where the flow slows and eddies form.
Collect water in the bench volumetric measuring tank for a timed period and divide the volume by the time to get the flow rate. Divide that by the cross-sectional area at each tapping for the mean velocity. The velocity head is the velocity squared divided by twice the acceleration due to gravity. At the throat, use the 25 mm diameter to work out the area.
Reversing the conduit changes which taper the water meets before the throat. Water speeding up in a narrowing taper follows the equation closely. Water slowing in a widening taper can separate from the walls and lose energy, more so when the steeper 21° taper is the widening one. Comparing both shows where the equation holds well.
Pressure differences grow with the square of the velocity, so a low, a middle and a high flow show the trend clearly. Repeating each setting also checks that the timed collections give consistent flow rates. Set each flow with the control valve at the module end and wait for steady levels.
To request a quotation, send the product name, code and quantity through our contact page or add the item to your enquiry cart.