The Analytical Balance Mass Set holds the reference masses used to counterpoise samples on a two-pan analytical or chemical balance. Its main pieces are chrome-plated brass and the smallest fractions are light alloy. A 10 mg rider and a pair of forceps sit with them in a polished case, and the set is listed as accuracy class B.
| Parameter | Detail |
|---|---|
| Type | Complete mass (weight) set for analytical beam balances |
| Main weights | Brass, chrome plated |
| Fractional weights | Light alloy |
| Rider | 10 mg |
| Handling tool | Forceps, included |
| Storage | Polished case |
| Accuracy class | Class B |
| Product code | SLE/LB/10 |
| Minimum order | 10 units |
| Dispatch | Usually within 7 to 10 working days |
| Manufacturer | ScienceLab (Jain Scientific Equipments Pvt Ltd), Ambala, India |
Weighing an empty crucible. Put a clean, cool crucible with lid on the left pan of the analytical balance and build the counter-mass on the right pan. Write down each piece as it goes on, then read the total a second time from the empty places in the case and check it again as each piece goes back.
Checking the set against itself. Place one large piece on a pan and balance it with smaller pieces whose nominal values add up to the same total. Swap the two groups between the pans and repeat. A difference that remains after the swap comes from the pieces rather than unequal balance arms, so those pieces should be compared with a reference set.
Water of crystallisation. Weigh a crucible, add hydrated copper sulfate and weigh again on a chemical balance. Heat gently until the blue crystals turn white, cool and weigh, then repeat to constant mass. The loss gives the water driven off, and the ratio of moles gives the formula of the hydrate.
Class B is the accuracy class the set is listed under. An accuracy class fixes how far each piece may differ from its nominal mass, and a more exacting class allows a smaller error. When a balance must be checked to a stated standard, use weights of the class that standard calls for.
No single interval fits every laboratory, because it depends on use and handling. Many departments compare the pieces with a reference set once a year, and sooner after a piece is dropped, scratched or corroded. Keep a written record of each check so slow drift becomes visible.
Fingerprints, dust, corrosive fumes, knocks and wear from sliding pieces across a pan all add or remove a little mass. Chrome plating protects the brass, but a scratch through it lets the metal tarnish. Keep the case closed, away from acid and ammonia fumes, and have any dented piece checked.
A fraction of a gram made in brass would be a tiny chip that is hard to pick up and easy to lose. Light alloy gives a larger piece for the same mass, so it can be gripped with the forceps and its marking read without difficulty.
The rider sits on the graduated beam and is moved with the rider slide. Its effect depends on where it sits: at the outer end of a scale graduated for this rider it counts as its full 10 mg, and nearer the centre it counts proportionally less. Read the value from the beam graduation it rests on.
Start with the heaviest piece that does not overshoot, then try each smaller piece in turn and keep those that do not tip the beam. Working down in strict order brings the balance close to equilibrium quickly with the fewest pieces. Finish with the light alloy fractions and then the rider.