The High Precision Balance weighs, counts and compares samples on a laboratory bench, using a VFD display and automatic external calibration. It offers counting and percentage modes, several weighing units, a built-in under hook for below-balance weighing and a standard RS 232 C output. Density measurement is an option, and capacity and readability are confirmed on enquiry.
| Parameter | Detail |
|---|---|
| Display | VFD (vacuum fluorescent display) |
| Calibration | Automatic external |
| Response time and tare range | 3–4 seconds; full tare, up to the maximum capacity of the balance |
| Weighing modes | Multifunction weighing units; counting; percentage (%) weighing |
| Below-balance weighing | Built-in under hook; density measurement optional |
| Data output | RS 232 C interface, standard |
| Operating temperature | 10°C to 40°C |
| Housing and weight | 330 x 200 x 100 mm; approx. 8.5 kg |
| Capacity and readability | Not stated in the listing; confirmed on enquiry |
| Product code | SLE/LB/15 |
| Minimum order | 10 units |
| Dispatch | Usually within 7 to 10 working days |
| Supplied by | ScienceLab (Jain Scientific Equipments Pvt Ltd), Ambala, India |
Uniformity of weight of tablets. Lift tablets one at a time with fine forceps into a tared weighing boat and record each mass. Students then find the mean and check every tablet against the limit their pharmacopoeia gives for that mean weight.
Counting small parts. Count a reference sample of identical screws, beads or capsules by hand, enter it in counting mode, and pour in the rest of the batch. The balance divides the total mass by the stored average piece weight and shows the count.
Density by suspension. With the balance set up for below-balance weighing as the supplied operating instructions describe, hang a solid from the under hook on cotton thread. Weigh it in air, then fully immersed in a tall form beaker of water. The loss in weight gives its volume and so its density.
Use a larger reference sample when the pieces differ slightly from one another. The balance multiplies any error in the stored average piece weight across the whole batch, so a bigger, carefully hand-counted sample gives a more reliable count.
Place a correctly filled reference item on the pan and store it as the 100 percent value. Each item weighed after that reads as a percentage of the reference, so overfilled or underfilled units stand out at once.
Yes. The standard RS 232 C interface connects to a printer or computer through a serial cable. Make the communication settings match on both ends, following the supplied operating instructions, and readings arrive without hand copying.
No. The under hook is built in, so a solid can be weighed in air and then hanging in water, and its density worked out by hand from the two readings. The density measurement function is a separate option; the under hook and normal weighing work with or without it.
Yes. The tare range is full, up to the maximum capacity of the balance, so even a heavy container can be zeroed out. The container still uses up capacity, though, so the sample can only weigh as much as the capacity minus the container.
Weigh each tablet and find the mean mass. Subtract the mean from each tablet's mass, divide by the mean and multiply by 100. Compare every result with the limit your pharmacopoeia sets for that mean weight.