The Lever Balance Dual Scale reads mass straight from a pointer, so no weights are added during a practical. One scale covers 0–250 g in 1 g steps and the other 0–1000 g in 5 g steps. The frame is die-cast aluminium alloy, zero is set with a levelling screw, and the detachable pan carries a hook for specific gravity work.
| Parameter | Detail |
|---|---|
| Type | Simple single-pan lever balance with two scales |
| Range 1 | 0–250 g x 1 g |
| Range 2 | 0–1000 g x 5 g |
| Frame | Die-cast aluminium alloy |
| Other parts | Chrome plated |
| Zero setting | Levelling screw |
| Reading | Pointer, read straight on to avoid parallax error |
| Pan | Detachable, with hook for specific gravity experiments |
| Product code | SLE/LB/12 |
| Minimum order | 10 units |
| Dispatch | Usually within 7 to 10 working days |
| Manufacturer | ScienceLab (Jain Scientific Equipments Pvt Ltd), Ambala, India |
Relative density of a solid. Tie a cube from the metal density cubes set to the hook with cotton thread and note its weight in air. Lower it fully into a glass beaker of water, clear of the sides and base, and note the smaller reading. Divide the weight in air by the loss in weight.
Upthrust in different liquids. Weigh one sinker in air, in water and then in strong salt solution held in a tall form beaker. The loss in the salt solution divided by the loss in water gives the relative density of that liquid.
Everyday masses for junior classes. Pupils weigh stones, fruit and metal blocks, reading the 250 g scale for light items and the 1000 g scale for heavier ones. Checking one reading against a known piece from a 250 g brass mass set shows how close the scale is.
Make sure nothing touches the pan or the lever and that the balance stands on a firm, flat bench. With the empty pan in place, turn the levelling screw until the pointer rests on zero. If it still wanders, look for dirt or a bent part fouling the lever.
Read the 0–250 g scale whenever the load is 250 g or less, because its 1 g divisions give the finer result. For heavier loads, read the 0–1000 g scale, where each division stands for 5 g.
Put a known mass on the pan and compare the pointer with its marked value at a few points on each scale. Remove it, set it back and read again. A sound balance repeats the same reading for the same load.
Water pushes up on the immersed solid with a force equal to the weight of water it displaces. The pointer drops by that amount, the apparent loss in weight, which also tells you the volume of the solid.
Air bubbles stuck to the solid, the solid touching the beaker, a thick or wet thread, a load still swinging and reading the pointer at an angle all give wrong values. Tap bubbles off, wait for the load to hang still and look straight at the pointer.
To request a quotation, send the product name, code and quantity through our contact page or add the item to your enquiry cart.
Last updated: 14 November 2026